11.4 Specifications
TRANSDUCER INPUT
Transducer type full-bridge, AC/AC-LVDT
Excitation Voltage 3 VRMS @ 5kHz,
short circuit protected
Transducer full-scale output .1 to 15 VRMS @ 3 VRMS excitation
Amplifier Gain Selection automatic
Calibration Type 2-, 3- or 5- point known displacement
A/D Converter 24-bit Sigma-Delta
Low-pass filter digital, 24-tap FIR
Resolution see page 83
Frequency Response see page 83
AUXILLIARY INPUTS
Quantity 2
Type momentary contact closure
Response Time < 5ms
Field-Selectable Functions tare on, tare off,
peak/valley clear, peak/valley hold,
track hold
ANALOG OUTPUT
Voltage range (voltage output
channels)
5, ±5, 10 or ±10 VDC
Current range (current output
channels)
4-20 mA
Source any channel’s track, peak or valley value
Isolation 500V
Resolution 13 bits
Frequency Response same as input if driven by the same chan-
nel’s tracking data