Honeywell UDA2182 Manual de usuario Pagina 153

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 258
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 152
Input Calibration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 141
8.5.2 Calibrating pH Electrodes Using Automatic Buffer recognition
Analyzer stores information on multiple buffers
The UDA2182 Universal Dual Analyzer contains (in its permanent memory) information
on several commonly used buffer solution standards in three groups, including the pH
versus temperature characteristics of each.
By command, the instrument will automatically select one of these buffers in the selected
group and use its values in the calibration process. Automatic checks are included to
ensure that reasonable and correct values are entered.
The procedure for using the automatic buffer recognition feature in an actual calibration is
provided in Table 8-2.
The standard pH Buffer values are listed in Table 8-1.
Vista de pagina 152
1 2 ... 148 149 150 151 152 153 154 155 156 157 158 ... 257 258

Comentarios a estos manuales

Sin comentarios