Honeywell UDA2182 Manual de usuario Pagina 155

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 258
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 154
Input Calibration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 143
Procedure
Make sure you have selected “PV Type –pH Glass, pH Durafet, or pH HPW” in the Inputs
configuration -Table 6-5.
Refer to Section 6.4.1 –
General Rules for Editing.
Table 8-2 Calibrating pH Electrodes Using Automatic Buffer Recognition
Step Action Screen
1
Prepare containers of two
standard reference solutions.
2
Press
Calibrate
CALIBRATION
Input Temp Cal
Output Cal
Cal History
Input PV Cal
CALIBRATION
Input Temp Cal
Output Cal
Cal History
Input PV Cal
Use
to select
Input PV Cal
3
Press
Enter
PV INPUT CAL
In 1 pH/ORP Cal
In 2 Conduc Cal
PV INPUT CAL
In 1 pH/ORP Cal
In 2 Conduc Cal
Use
to select
Input 1 or 2 pH/ORP Cal
4
Press
Enter
IN 1 pH/ORP Cal
Sample Cal
Auto Buffer Cal
Buffer Group
pH Offset
pH Slope
Reset pH Offset
Reset pH Slope
Buffer Cal
IN 1 pH/ORP Cal
Sample Cal
Auto Buffer Cal
Buffer Group
pH Offset
pH Slope
Reset pH Offset
Reset pH Slope
Buffer Cal
Use
to select
”Buffer Group”
5
Press
Enter
Use
to select
NIST/USP (default)
USA, or
Europe
Vista de pagina 154
1 2 ... 150 151 152 153 154 155 156 157 158 159 160 ... 257 258

Comentarios a estos manuales

Sin comentarios